Generator Specifications:
- KVP 1 KV steps, 40 to 125 KVP

- Milliamp Five position switch; 60S, 120S,
240L, 300L, 360L @60 Hz.
(Model; 2136) 50S, 100S, 150L, 200L,
300L @50 Hz. (Model; 2130, 2136)

- Timer Digital 23 position solid state,
1/120 (1 pulse) to 2 seconds with 60 Hz. Line.
Digital 23
position solid state, 1/100 (1 pulse) to 5 seconds with 50 Hz.
Line.
Contactors
SCRs with zero start & stop using
electromagnetic contactor as back up.
Tube Safety
Built in calibrateable circuit to protect
against Tube overload (KV, mA & Time)
Circuit Breaker
Mid-secondary breaker standard.
X-Ray Tube Filament
Constant voltage stabilizer working into a
temperature compensated calibrating network.
H.T. Tank
Single tube, full wave, silicon rectifiers
(Model; 2125)

Power
Requirements:
Power Requirement 200 to 250 Volt single
phase line rated at 100 amps with a regulation of at least 2%.
Maximum line current is 140 amps @ 230 Volts. Maximum line
resistance to enable compliance at full load is 0.15 Ohms. Duty
cycle is 1% or 300 mA @ 125 KVP every 10 minutesr frequency for
reduced radiation exposure.